神戸大学附属図書館デジタルアーカイブ
入力補助
English
カテゴリ
学内刊行物
ランキング
アクセスランキング
ダウンロードランキング
https://hdl.handle.net/20.500.14094/0100490459
このアイテムのアクセス数:
66
件
(
2025-05-07
12:41 集計
)
閲覧可能ファイル
ファイル
フォーマット
サイズ
閲覧回数
説明
0100490459 (fulltext)
pdf
1.55 MB
63
メタデータ
ファイル出力
メタデータID
0100490459
アクセス権
open access
出版タイプ
Version of Record
タイトル
The Utility and Feasibility of Smart Glasses in Spine Surgery: Minimizing Radiation Exposure During Percutaneous Pedicle Screw Insertion
著者
Hiranaka, Yoshiaki ; Takeoka, Yoshiki ; Yurube, Takashi ; Tsujimoto, Takeru ; Kanda, Yutaro ; Miyazaki, Kunihiko ; Ohnishi, Hiroki ; Matsuo, Tomoya ; Ryu, Masao ; Kumagai, Naotoshi ; Kuroshima, Kohei ; Kuroda, Ryosuke ; Kakutani, Kenichiro
著者名
Hiranaka, Yoshiaki
著者ID
A3390
研究者ID
1000030943592
KUID
https://kuid-rm-web.ofc.kobe-u.ac.jp/search/detail.html?systemId=4a9f37836e6eafb2520e17560c007669
著者名
Takeoka, Yoshiki
武岡, 由樹
タケオカ, ヨシキ
所属機関名
医学研究科
著者ID
A2131
研究者ID
1000010514648
KUID
https://kuid-rm-web.ofc.kobe-u.ac.jp/search/detail.html?systemId=564b5b743f5a7d84520e17560c007669
著者名
Yurube, Takashi
由留部, 崇
ユルベ, タカシ
所属機関名
医学部附属病院
著者名
Tsujimoto, Takeru
著者名
Kanda, Yutaro
著者名
Miyazaki, Kunihiko
著者名
Ohnishi, Hiroki
著者名
Matsuo, Tomoya
著者名
Ryu, Masao
著者名
Kumagai, Naotoshi
著者名
Kuroshima, Kohei
著者ID
A0783
研究者ID
1000080379362
KUID
https://kuid-rm-web.ofc.kobe-u.ac.jp/search/detail.html?systemId=f51cbfc82378ae63520e17560c007669
著者名
Kuroda, Ryosuke
黒田, 良祐
クロダ, リョウスケ
所属機関名
医学部附属病院
著者ID
A1461
研究者ID
1000010533739
KUID
https://kuid-rm-web.ofc.kobe-u.ac.jp/search/detail.html?systemId=3401110282b6c620520e17560c007669
著者名
Kakutani, Kenichiro
角谷, 賢一朗
カクタニ, ケンイチロウ
所属機関名
医学研究科
言語
English (英語)
収録物名
Neurospine
巻(号)
21(2)
ページ
432-439
出版者
Korean Spinal Neurosurgery Society
刊行日
2024-06
公開日
2024-07-24
抄録
Objective: Spine surgeons are often at risk of radiation exposure due to intraoperative fluoroscopy, leading to health concerns such as carcinogenesis. This is due to the increasing use of percutaneous pedicle screw (PPS) in spinal surgeries, resulting from the widespread adoption of minimally invasive spine stabilization. This study aimed to elucidate the effectiveness of smart glasses (SG) in PPS insertion under fluoroscopy. Methods: SG were used as an alternative screen for fluoroscopic images. Operators A (2-year experience in spine surgery) and B (9-year experience) inserted the PPS into the bilateral L1–5 pedicles of the lumbar model bone under fluoroscopic guidance, repeating this procedure twice with and without SG (groups SG and N-SG, respectively). Each vertebral body’s insertion time, radiation dose, and radiation exposure time were measured, and the deviation in screw trajectories was evaluated. Results: The groups SG and N-SG showed no significant difference in insertion time for the overall procedure and each operator. However, group SG had a significantly shorter radiation exposure time than group N-SG for the overall procedure (109.1 ± 43.5 seconds vs. 150.9 ± 38.7 seconds; p = 0.003) and operator A (100.0 ± 29.0 seconds vs. 157.9 ± 42.8 seconds; p = 0.003). The radiation dose was also significantly lower in group SG than in group N-SG for the overall procedure (1.3 ± 0.6 mGy vs. 1.7 ± 0.5 mGy; p = 0.023) and operator A (1.2 ± 0.4 mGy vs. 1.8 ± 0.5 mGy; p = 0.013). The 2 groups showed no significant difference in screw deviation. Conclusion: The application of SG in fluoroscopic imaging for PPS insertion holds potential as a useful method for reducing radiation exposure.
キーワード
Smart glasses
Pedicle screw
Fluoroscopy
Augmented reality
Radiation exposure
カテゴリ
医学研究科
医学部附属病院
学術雑誌論文
権利
© 2024 by the Korean Spinal Neurosurgery Society
This is an open access article distributed under the terms of the Creative Commons Attribution Non-Commercial License which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
関連情報
DOI
https://doi.org/10.14245/ns.2448090.045
PMID
38955520
詳細を表示
資源タイプ
journal article
ISSN
2586-6583
OPACで所蔵を検索
CiNiiで学外所蔵を検索
eISSN
2586-6591
OPACで所蔵を検索
CiNiiで学外所蔵を検索
ホームへ戻る